Trade-World / Medical and Health / Medical Equipment
Multi-parameter Patient Monitor BD9000
Previous   Next

Multi-parameter Patient Monitor BD9000



12-channel Display of Drug Concentrate


with such functions as simultaneous acquisition of 7-channel and 12-channle ECG waveforms displayed on one screen, ST section analysis available, it is more convenient for clinical; three heart rate sources upon choice; waveform filling, 5-minute heart rate replay, able to measure multiple standard parameter and pulse transmit time, cuff blood pressure; complete patient information storage and monitor review functions.


 


Trend Review


With the link formed by network, bed-side machines and central multi-parameter monitor system, thus a monitor network is formed, so as to help the clinical staff to have real-time monitor; trend data display and 256h trend review.


 


Fetal Monitor  


Standard 6-parameter monitor, additional single and double invasive blood pressure; end-expiratory CO2 (by-flow); fetal monitor function able to precise fetal action and uterine contraction


 


Drug Concentrate Calculation


The powerful assistant to clinical staff, able to have drug calculation: drug concentrate, transfusion speed, drug amount etc provided by burette; 10 common drug provided by the database, and 5 drug customized.


 


Super-big Screen


12.1¡¯ TFT big-screen, real-color, wide-vision-angle and high-brightness LCD display, with Chinese-English interface


 


Removable Rechargeable Battery


Removable rechargeable battery, free of maintenance, able to work more than 2 hours


 


Printing and Interface


Manual printing, alarm trigger printing, real-time printing record and real-time abnormality monitor record: able to use together with high-frequency electrical cutter and defibrillator, high interference resistance.


 



[Contact Information]

Posted by John Liu
e-MailThis information shows after Login.
Firm
JETEEM MEDICAL ELECTRONICS TECH CO.,LTD.
Address shenzhen, Guang dong province, China
City & ZIP
518055
Land
China
Telephone
This information shows after Login.

InterestPatient Monitor

Product Specifications
Technical Specifications
Size and weight
Size: 320mm¡Á270mm¡Á155mm£¨BD9000£©
345mm¡Á260mm¡Á168mm (BD6000)
Weight: 7.5 KG
Operating condition
Temperature range: Operating 0-45¡æ
Saving and transporting -20-60¡æ
Dampness range: Operating <=85%
Saving and transporting <=95%
Altitude range: Operating -500-4,600m
Saving and transporting -500-13,100m
Electrical specification: AC 100-250V 50/60 Hz, Max input power 60VA,Fuse T 0.5A
Specification of screen
True color,12.1¡±TFT type LCD, distinguish rate 800¡Á600
Signal interface
Interface of exterior display: Standard VGA
Exterior communication interface: RS422/RS485
Exterior printer interface: Standard combines
Record apparatus
Record width: 50mm
Paper feed speed: 25mm/s
Recording waveform: 3 ways
Record type: continuously real time record
Manually one time record
Time striking record
Alarm striking record
Trend saving
6 hours/1 second¡ª128 hours/1 minute
Specification of ECG
Lead mode: 12 leads (R,L,F,N,C1¡¢C2¡¢C3¡¢C4¡¢C5¡¢C6)
Method of Lead: ¢ñ¡¢¢ò¡¢¢ó¡¢aVR¡¢aVL¡¢aVF¡¢V1¡¢V2¡¢V3¡¢V4¡¢V5¡¢V6
Waveform of Lead: 12 ways
Gain: ¡Á0.5 ¡¢¡Á1¡¢¡Á2¡¢¡Á4
Mode: diagnoses mode, monitoring mode, and operating mode
Input resistance: >5 M¦¸
Coenelasma suppression ratio: >90dB
Polarization voltage range of electrode: ¡À300mV
Drain current: <30¦ÌA
Resuming time of base line: after defibrillation<3s
Range of ECG signal: ¡À10 mV (peak value)
Frequency response: Diagnoses mode 0.05-100Hz
Monitoring mode 0.5-40Hz
Operation mode 1-20Hz
Heart rate
Measuring range: 10-300 bmp
Measuring accuracy: ¡À1%
Definition: 1bpm
Sensitivity ratio: >0.2mV (peak value)
Respiration specification
Method: RA-LA resistance
Input the resistance: >2.5 M¦¸
Bandwidth: 0.1-2Hz
Respiration rate: 0-120 rpm
Measuring range: 0-120 rpm
Definition: 1 rpm
Sensitivity ratio: ¡À2 rpm
Specification of NIBP
Measuring method: method of shocking automatically
Operating mode: manual, periodic and continuous
Partition of periodic measure: 5,10,45,60 minutes
Continuous mode: 5 minutes
Range of pulse rate: 40-240 bpm
Measuring range:
Adult: Systolic pressure 25-255 mmHg
Diastolic pressure 10-220 mmHg
Mean pressure 15-235 mmHg
Child: Systolic pressure 20-135 mmHg
Diastolic pressure 10-110 mmHg
Mean pressure 15-125 mmHg
Definition: 1 mmHg
Sensitivity ratio: Max average error ¡À5 mmHg
Max standard error 8 mmHg
Overvoltage protection: Adult 290 mmHg
Child 145 mmHg
Specification of SPO2
Measuring range: 0-100%
Definition: 1%
Sensitivity ratio: 70-100% ¡À2%
40-69% ¡À3%
0-39% not explain
Refresh time: About a second
Alarm delay: 10 seconds
Pulse rate: Measuring range 0-254 bmp
Definition 1bmp
Sensitivity ratio ¡À2 bpm
Alarm delay 10 seconds
Specification of TEMP
Measuring range: 0¡æ-50¡æ
Definition 0.1¡æ
Sensitivity ratio: ¡À0.2¡æ
Refresh time: About a second
Average time constant: <10s



Industry Categorymedical equipment

Quality/Safety Certifications

Delivery Lead Timewithin 20 days

Payment & Price TermsFOB T/T

My Offerswelcome you to ask about it

Other ProductsPatient Monitor
Fetal Monitor
Fetal Doppler
Pulse Oximeter

Customersall over the world

Company History

Estimated Annual Sales

Company News

Remarks

Web SiteThis information shows after Login.

Product Details
Product id 26762
views  158
Added 2007.04.20
 

content

Products in this categoryMore productsContact nowSend to friendRate itPrint